Messages and Diagnostic Codes
Page 72 34-ST-25-20 MC Toolkit User Manual Release 0
7/03
MC TOOLKIT ERROR MESSAGES
Memory Error If several programs are active, try closing one or more open
programs.
Too many programs installed in the Pocket PC. Check the Pocket
PC free memory. Uninstall programs if need to release memory.
Too many data files in the Pocket PC. Check the Pocket PC free
memory. Delete unused data files to release memory.
Use the Pocket PC hardware reset. See your Pocket PC
documentation reset the unit. Caution: If the Symbol PPT 2800
computer is reset, it will lose all its installed programs and data. The
program will have to be installed again.
Message length is longer than expected
Message length is shorter than expected
Communications problem between the Pocket PC computer and the
modem. Repeat the command again.
Modem bad checksum
This could indicate a defective modem. Repeat the command again.
If the problem persist contact Honeywell TAC.
Modem Buffer overflow
Modem busy
Modem framing error
Modem illegal command
Modem illegal data
Modem network error
Modem Transmitter Serial Error!
Communications problem between the Pocket PC computer and the
modem. Repeat the command again.
No Response from the Transmitter
Make sure that the MC Toolkit field connections are connected to the
transmitter.
Verify the transmitter is wired correctly and that it is powered.
If connected to a DE transmitter, make sure the polarity of the cables
connecting to the transmitter is correct.
If connected to a HART transmitter, make sure the address number is
correct.
Make sure the correct protocol for the transmitter is selected.
A defective transmitter can also cause this error message.
Verify that a 250 ohm resistor in series with the transmitter.
Resume Monitor Thread Failed! Close all the other running applications.
Restart MC Toolkit application.
Vista de página 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 109 110

Comentários a estes Manuais

Sem comentários